Skip to content

PlusServer connection to NDI Polaris tracker with Data averaging activated fails #1095

Description

@orubioDeneb

Since Firmware version 008, NDI Polaris Vega trackers allow to configure averaged data output by dividing its tracking frequency by a factor.

When configuring the tracker with Data averaging factor 2 with tracking frequency 400Hz, and trying to connect with PlusServer, a bad CRC error message arrives to the log constantly when trying to start the server.

Changing tracking configuration file to 200Hz is also not working. It is probably trying to change the tracking frequency to 200Hz and then performing the data averaging at that frecuency, but couldn't trace this more at this time.

NDI Track app is correctly working with this setup, so it does not seem a configuration error. Using FW revision 010 at this moment but same error appeared with any other version tested since revision 008. Trying to force tracking output to 1/n factor in NDI Configuration App also did not work.

Could be related to pull request #1068 as capture and tracking frequency differ when this option is activated?

Thank you very much

Metadata

Metadata

Assignees

No one assigned

    Labels

    No labels
    No labels

    Type

    No type

    Projects

    No projects

    Milestone

    No milestone

    Relationships

    None yet

    Development

    No branches or pull requests

    Issue actions