DTSA-II and EDS analysis of the first NAB (C63000) powder: SEM/EDS of Oct 27, 2026 (agent-staging 3c25afa) #27

Open

Description

@sgbaird

@claude Run DTSA-II and any other EDS analysis you think worthwhile on the first NAB powder from our atomizer. Sterling asked for this on Oct 28: "run whatever DTSA-II or other kinds of analysis you might so be inclined to on that."

Data

The data are staged in sgbaird/agent-staging, folder for/vertical-cloud-lab/caliber/2026-10-28-nab-c63000-sem-eds/, commit 3c25afa. Read them with GH_READ_PAT_STAGING, and check every file against the sha256 list in MANIFEST.md. The folder has a .gitattributes that keeps the CRLF text files byte-exact.

  • Spectra: three spectra of the same powder at 10, 15 and 20 kV and 3.2 nA, EDAX Octane Plus, in .msa, .spc, .csv and .xml formats. The 20 kV one is the one Ronnie marked "USE THIS ONE": 24.4 s live time, take-off angle 36°, 5 eV per channel.
  • SEM images: four original Apreo TIFFs, with the FEI metadata in TIFF tag 34682.
  • Sample: wrought 630 bronze (UNS C63000, a nickel-aluminum bronze, nominally about 82Cu-10Al-5Ni-3Fe), ultrasonically atomized on Oct 27 into spherical powder, mostly about 30 to 180 µm.
  • Mounting: on carbon tape. The stub material is not recorded.
  • No quantification was exported, so there are no TEAM wt% values to compare against.

What I'd like

Use your judgment and do more if it's useful. At minimum:

  1. Peak identification check. TEAM found C, Al, Fe, Ni and Cu. Look for anything else: W from the tungsten-dipped plate (W M near 1.78 keV, next to Al Kα), O, Mn (C63000 allows up to 1.5%), Si, and escape or sum peaks.

  2. Quantification with DTSA-II (NIST; headless with its Jython scripting if you can). Use standardless or simulated standards, at all three voltages, and compare against the C63000 nominal and spec ranges. Report the uncertainties, and say plainly what the rough, spherical, 30 to 180 µm particles on carbon do to accuracy. A flat, polished section would differ.

  3. Monte Carlo simulation in DTSA-II of the nominal C63000 composition at 10, 15 and 20 kV:

    • as a bulk flat sample;
    • as a sphere of about 50 to 100 µm on a carbon substrate.

    Compare the simulated and measured spectra, especially the low-energy L lines against the K lines.

  4. Acquisition advice in CALIBER's terms. Which voltage, current and live time would give quantification good to about ±1 wt% for Al, Ni and Fe in this alloy? Consider overvoltage for Cu K (edge 8.98 keV), dead time, and the counts needed for Fe and Ni at 3 to 5 wt%. Should the next session use polished cross-sections? Also recommend a measurement protocol for the next powder lots.

  5. Cross-check with a second tool if it's cheap, for example HyperSpy/exspy for peak fitting or the NIST NeXL tools.

Constraints

  • Where to put it. Commit the analysis under a new folder such as analysis/2026-10-28-nab-c63000-eds/: scripts, outputs, figures, and a README of the results with the staging commit as provenance. Open a PR into main, and don't merge it.
  • Push early. Push after each step. At the first 401, or by minute 50, run python scripts/refresh_github_app_token.py and keep going.
  • Scope. Don't write to agent-staging, and don't touch other repos.
  • Final comment. Include a short results table (measured and simulated wt% against nominal), the main figure embedded by commit-hash URL, and the acquisition recommendation.

Activity

  1. claude commented on Oct 28, 2026

    @claude

    Claude finished @sgbaird's task in 50m 2s —— View job • claude/issue-27-20261028-0415 • [Create PR ➔](https://github.com/vertical-cloud-lab/caliber/compare/main...claude/issue-27-20261028-0415?quick_pull=1&title=Issue%20%2327%3A%20Changes%20from%20Claude&body=This%20PR%20addresses%20issue%20%2327%0A%0AGenerated%20with%20%5BClaude%20Code%5D(https%3A%2F%2Fclaude.ai%2Fcode)


    NAB (C63000) powder: DTSA-II and EDS analysis

    I quantified all three spectra in DTSA-II and simulated them with its Monte Carlo. At 20 kV the powder is inside the C63000 limits for Al, Fe, Ni and Mn, and no tungsten from the plate is detected. Everything is in analysis/2026-10-28-nab-c63000-eds/ on PR #28 (not merged). All 20 staged files matched their sha256 in MANIFEST.md. I didn't commit the raw spectra or SEM images; fetch_data.sh re-fetches and checks them.

    Results (wt%, normalized to 100% metals, ±1σ):

    Al Fe Ni Mn Cu
    C63000 limits 9.0–11.0 2.0–4.0 4.0–5.5 ≤1.5 bal.
    Nominal (the Monte Carlo input) 10 3 5 – 82
    Measured 20 kV, DTSA-II 9.2 ± 2.4 3.42 ± 0.11 5.33 ± 0.17 0.14 ± 0.03 81.9 ± 2.2
    Measured 15 kV, DTSA-II 8.2 ± 1.5 3.33 ± 0.08 5.10 ± 0.13 0.08 ± 0.04 83.3 ± 1.4
    Measured 10 kV, DTSA-II 6.0 ± 0.7 2.77 ± 0.13 4.43 ± 0.33 n.d. 86.8 ± 0.7
    Measured 20 kV, exspy + Monte Carlo as standard 9.7–10.1 3.55–3.81 4.91–4.97 – 81.1–81.8
    Simulated flat bulk, 20 kV, run back through DTSA-II 10.03 2.91 4.97 – 82.1
    Simulated raster over a 50 µm sphere on C, 20 / 15 / 10 kV 9.65 / 8.71 / 7.44 2.72 / 2.76 / 2.83 4.87 / 4.93 / 4.86 – 82.8 / 83.6 / 84.9

    Peak identification

    • No tungsten: W Mα fits to zero, so W is below about 0.2 wt% (3σ), and the clean W Lβ1 line at 9.67 keV is absent.
    • The bump next to Al K is a pile-up peak, not W or Si. It sits at 1.82 keV (two Cu L-region photons) and grows with the square of the count rate. A Cu Kα + Cu Kα pile-up peak also shows at 16.1 keV.
    • Trace and light elements: Mn is present at about 0.1 wt%. Si (below 0.1 wt%) and Zn (below 0.25 wt%) were not detected. O is present (surface oxide), and the large C peak is the tape. Escape peaks are negligible, at most 0.05 wt% on Fe.
    • Energy scale: it was about 10 eV high at Cu Kα, which I corrected per spectrum. Before that fix, a first pass read 2.3 wt% Mn.

    What the loose spheres on carbon do to accuracy

    • On a particle top it is effectively bulk. The 50 and 100 µm sphere simulations match flat bulk within a few percent.
    • An area scan over particles is not. Shadowed and tilted surfaces absorb Al K and the L lines much more than the K lines. In the simulation Al reads 0.4, 1.1 and 1.8 wt% low at 20, 15 and 10 kV, and the measured Al K / Cu Kα ratio follows that trend within 3–6% at 15–20 kV. That is the most likely reason Al drops to 6 wt% at 10 kV.
    • Analytical totals: the measured spectra give 30–39%, against 65–70% for the simulated raster and about 100% for flat bulk.
    • The L lines don't agree with the K lines (Cu L / Cu Kα comes out 1.27× the simulation, Ni L / Ni Kα 0.6–0.8×), so I didn't use them for this alloy.
    • Accuracy without standards: Newbury (1998) found 95% of standards-based analyses on flat samples within ±5% relative, against ±25–50% without standards.

    Acquisition advice (in CALIBER's parameters: beam energy, current, acquisition time)

    • Sample: use polished cross-sections for quantification, and keep loose-powder stubs for morphology only.
    • Beam energy: 20 kV, with 15 kV as a second setting. Overvoltage for Cu K is 2.2 at 20 kV but only 1.1 at 10 kV, so drop 5 and 10 kV for quantification.
    • Current: 1.3–2.0 nA at 20 kV, for 20–30% dead time. The 3.2 nA used here gave 42 kcps out (about 43% dead time) and visible pile-up. Measure the current with a Faraday cup and save the real time.
    • Live time: 100–200 s. Counting statistics are not the limit: even the existing 24 s at 20 kV gives Fe ±0.04, Ni ±0.07 and Al ±0.09 wt%. Measure at least 10 locations per lot.
    • ±1 wt% on Fe and Ni is already met.
    • ±1 wt% on Al needs measured standards (Cu, Al, Ni, Fe, Mn, plus a polished coupon of the feedstock rod) on a flat sample. Without standards, DTSA-II's own Al uncertainty is ±1.5–2.4 wt%.
    • Reference: send the rod and each lot for ICP-OES/MS; that is the (EDX − ICP-MS)² objective.
    • The full protocol is in §4 of the README.

    Limits

    • I assumed the detector area (30 mm²) and window (Moxtek AP3.3 polymer); the .spc header has neither. They affect Al and the L lines, not the K-line ratios.
    • The analysed area isn't recorded. The SEM frames are 5 kV images, and the large C peak suggests the spectra were area scans that included tape.
    • I used HyperSpy/exspy for the second tool and did not try NIST NeXL.

    Commands (all exited 0)

    ./fetch_data.sh                                  # 20/20 sha256 OK
    python eds_peaks.py /tmp/nab-staging/files/eds   # calibration, peak checks
    ./run_dtsa2.sh dtsa2_quant.py                    # DTSA-II Polaris, Xvfb + Java 26 + --enable-preview
    ./run_dtsa2.sh dtsa2_mc.py                       # 12 MC spectra, about 12 min
    ./run_dtsa2.sh dtsa2_quant.py                    # measured + MC spectra
    python exspy_crosscheck.py ...; python compare_mc.py ...
    

    · Branch · PR #28

  2. sgbaird commented on Oct 28, 2026

    @sgbaird
    MemberAuthor

    Correction to the staged MANIFEST.md (for the record; no action needed): sem/3.2nA.tif is a 4,379x close-up (HFW 94.6 µm, 0.0308 µm/px, 5 kV, 3.2 nA, 18:05), not a 150x field. sem/3.2nA_002.tif is the 150x field (HFW 2.76 mm, 0.899 µm/px). The spectra were acquired at 18:19–18:24, after both. Per-image metadata for all 16 images of the session is in the DAL drop, for/vertical-cloud-lab/digital-alloy-lab/2026-10-28-nab-c63000-sem-psd/MANIFEST.md (agent-staging 485dd7b).

Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment

Metadata

Metadata

Assignees

No one assigned

    Labels

    No labels
    No labels

    Type

    No type

    Projects

    No projects

      Milestone

      No milestone

      Relationships

      None yet

      Development

      No branches or pull requests

      Issue actions