Repository navigation
Description
@claude Run DTSA-II and any other EDS analysis you think worthwhile on the first NAB powder from our atomizer. Sterling asked for this on Oct 28: "run whatever DTSA-II or other kinds of analysis you might so be inclined to on that."
Data
The data are staged in sgbaird/agent-staging, folder for/vertical-cloud-lab/caliber/2026-10-28-nab-c63000-sem-eds/, commit 3c25afa. Read them with GH_READ_PAT_STAGING, and check every file against the sha256 list in MANIFEST.md. The folder has a .gitattributes that keeps the CRLF text files byte-exact.
- Spectra: three spectra of the same powder at 10, 15 and 20 kV and 3.2 nA, EDAX Octane Plus, in
.msa,.spc,.csvand.xmlformats. The 20 kV one is the one Ronnie marked "USE THIS ONE": 24.4 s live time, take-off angle 36°, 5 eV per channel. - SEM images: four original Apreo TIFFs, with the FEI metadata in TIFF tag 34682.
- Sample: wrought 630 bronze (UNS C63000, a nickel-aluminum bronze, nominally about 82Cu-10Al-5Ni-3Fe), ultrasonically atomized on Oct 27 into spherical powder, mostly about 30 to 180 µm.
- Mounting: on carbon tape. The stub material is not recorded.
- No quantification was exported, so there are no TEAM wt% values to compare against.
What I'd like
Use your judgment and do more if it's useful. At minimum:
-
Peak identification check. TEAM found C, Al, Fe, Ni and Cu. Look for anything else: W from the tungsten-dipped plate (W M near 1.78 keV, next to Al Kα), O, Mn (C63000 allows up to 1.5%), Si, and escape or sum peaks.
-
Quantification with DTSA-II (NIST; headless with its Jython scripting if you can). Use standardless or simulated standards, at all three voltages, and compare against the C63000 nominal and spec ranges. Report the uncertainties, and say plainly what the rough, spherical, 30 to 180 µm particles on carbon do to accuracy. A flat, polished section would differ.
-
Monte Carlo simulation in DTSA-II of the nominal C63000 composition at 10, 15 and 20 kV:
- as a bulk flat sample;
- as a sphere of about 50 to 100 µm on a carbon substrate.
Compare the simulated and measured spectra, especially the low-energy L lines against the K lines.
-
Acquisition advice in CALIBER's terms. Which voltage, current and live time would give quantification good to about ±1 wt% for Al, Ni and Fe in this alloy? Consider overvoltage for Cu K (edge 8.98 keV), dead time, and the counts needed for Fe and Ni at 3 to 5 wt%. Should the next session use polished cross-sections? Also recommend a measurement protocol for the next powder lots.
-
Cross-check with a second tool if it's cheap, for example HyperSpy/exspy for peak fitting or the NIST NeXL tools.
Constraints
- Where to put it. Commit the analysis under a new folder such as
analysis/2026-10-28-nab-c63000-eds/: scripts, outputs, figures, and a README of the results with the staging commit as provenance. Open a PR intomain, and don't merge it. - Push early. Push after each step. At the first 401, or by minute 50, run
python scripts/refresh_github_app_token.pyand keep going. - Scope. Don't write to
agent-staging, and don't touch other repos. - Final comment. Include a short results table (measured and simulated wt% against nominal), the main figure embedded by commit-hash URL, and the acquisition recommendation.
Activity
claude commentedon Oct 28, 2026
Claude finished @sgbaird's task in 50m 2s —— View job •
claude/issue-27-20261028-0415• [Create PR ➔](https://github.com/vertical-cloud-lab/caliber/compare/main...claude/issue-27-20261028-0415?quick_pull=1&title=Issue%20%2327%3A%20Changes%20from%20Claude&body=This%20PR%20addresses%20issue%20%2327%0A%0AGenerated%20with%20%5BClaude%20Code%5D(https%3A%2F%2Fclaude.ai%2Fcode)NAB (C63000) powder: DTSA-II and EDS analysis
I quantified all three spectra in DTSA-II and simulated them with its Monte Carlo. At 20 kV the powder is inside the C63000 limits for Al, Fe, Ni and Mn, and no tungsten from the plate is detected. Everything is in
analysis/2026-10-28-nab-c63000-eds/on PR #28 (not merged). All 20 staged files matched their sha256 inMANIFEST.md. I didn't commit the raw spectra or SEM images;fetch_data.shre-fetches and checks them.Results (wt%, normalized to 100% metals, ±1σ):
Peak identification
What the loose spheres on carbon do to accuracy
Acquisition advice (in CALIBER's parameters: beam energy, current, acquisition time)
Limits
.spcheader has neither. They affect Al and the L lines, not the K-line ratios.Commands (all exited 0)
· Branch · PR #28
sgbaird commentedon Oct 28, 2026
Correction to the staged
MANIFEST.md(for the record; no action needed):sem/3.2nA.tifis a 4,379x close-up (HFW 94.6 µm, 0.0308 µm/px, 5 kV, 3.2 nA, 18:05), not a 150x field.sem/3.2nA_002.tifis the 150x field (HFW 2.76 mm, 0.899 µm/px). The spectra were acquired at 18:19–18:24, after both. Per-image metadata for all 16 images of the session is in the DAL drop,for/vertical-cloud-lab/digital-alloy-lab/2026-10-28-nab-c63000-sem-psd/MANIFEST.md(agent-staging485dd7b).